Bellingham + Stanley Polarimeter ADP640 with 405nm, 436nm, 546nm & 589nm Multiple Wavelengths and Peltier Temperature Control

Bellingham + Stanley Polarimeter ADP640 with 405nm, 436nm, 546nm & 589nm Multiple Wavelengths and Peltier Temperature Control

  • SKU: 37-64
Request a Quote
  • Description

    Bellingham + Stanley Polarimeter ADP640 with 405nm, 436nm, 546nm & 589nm multiple wavelengths and peltier temperature control for measuring optical rotation. Suited for use by scientists wishing to measure chiral compounds and other optically active substances in the chemical, pharmaceutical and food sectors as well as for use in academic research.

    • Peltier temperature controlled
    • Four decimal place resolution
    • Conforms to USP/EP/BP/JP
    • RFID user clearance
    • High Definition touchscreen
  • Specifications
    General Specifications 
    Range (°A) ± 89 (-355 to +355 via Method selection)
    Resolution (°A) 0.0001
    Accuracy (°A) ± 0.003 (@546 & 589nm) / ± 0.005 (@325, 365, 405, 436 & 633nm)
    Temperature Range 15-35°C
    Temperature Control / Accuracy Peltier / ± 0.2°C
    Temperature Compensation None, sugar, quartz, user defined
    Optical Density Range 0.0 to 3.0 OD
    Methods Specific Rotation, % Concentration, % Invert Sugar, % Inversion (A-B)
    Temperature Set Points 20 & 25 °C (variable between 20-30 °C via Method)
    Reading Time 15-60 seconds @ 546/589nm and 20/20°C (instrument/sample)
    Tube Length 5-200mm
    Tube Diameter 3-8mm
    User Interface High Definition 7.4” touch-screen color display
    Light Source UV/Vis lamp (6V, 2A >1000hrs) and narrow band pass filter(s)
    Interfaces 4 x USB, 1 x Ethernet, 1 x RS232
    Power Supply 100-250V~, 50-60 Hz. <6A.