ISE sensor für nitrate - WTW

ISE sensor für nitrate - WTW

  • SKU: 107082
Request a Quote
  • Description
    Digital ISE sensor for nitrate, connectable to the IQ SENSOR NET
    Nitrogen elimination – transparent, process optimized, economical. Nitrate measurement directly in the medium - optimized for regulation purposes:
    •Chloride compensation (optional)
    •Calibration free, long-term stability
    •Economical & cost efficient
    With the NitraLyt®Plus 700 IQ, nitrate can be determined directly in the medium. This makes measured values available in a timely manner and they can easily be used for regulation purposes. Another plus of the IQ SENSOR NET probe is the low investment and maintenance cost.
  • Specifications

    Measuring method

    Electrochemical

    Appropriate Electrode

    Reference electrode VARiON® Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl

    Measuring range/Resolution

    NO3-N: 1 ... 1000 mg/l / 1 mg/l; 0.1 ... 100.0 mg/l / 0.1 mg/l
    NO3-: 5 ... 4500 mg/l / 5 mg/l; 0.5 ... 450.0 mg/l / 0.5 mg/l

    Compensation Range

    Cl-: 0.1 ... 1,000 mg/l / 0.1 mg/l

    Measuring Accuracy in laboratory standard solutions

    ± 5 % of measured value ± 0.2 mg/l in standard solutions

    Calibration Procedures

    Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution

    Working Life (typically)

    Reference electrode: 18 months,
    measuring and compensation electrode: 18 months (in typical application - municipal sewage plants)

    Temperature Measurement and Compensation

    Integrated NTC thermistor, 
    Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s

    Pressure Resistance

    Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes)

    Ambient Conditions

    Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C)

    Electrical connections

    2-wire shield cable with quick fastener to sensor

    Electromagnetic Compatibility

    EN 61326, Class B, FCC Class A;
    Intended for indispensable operation

    Certifications

    CE

    Mechanical

    Sensor body: V4A stainless steel 1.4571 Protective cup: POM
    Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes)
    Electrode connector: POM

    Weight

    Approx. 1.48 lb (670 g, without electrode, without sensor connection cable)

    Warranty

    NitraLyt®Plus 700 IQ: 2 years
    Electrodes: 1 year for defects of quality