Model
LRD 01
LRD 325
LR 325/01
LR 325/001
TetraCon® 325
TetraCon® DU/T
Order No.
302220, 302222
302225, 302229
301961
301962
301960, 301970, 301971
301252
Measuring principle
2 electrodes measuring cell
4 electrodes measuring cell
Measuring Range
0.001 µS/cm ... 200 µS/cm
1 µS/cm ... 2 S/cm
0.0001 µS/cm ... 30 µS/cm
Cell constant
0.1 cm-1, ±2%
0.475 cm-1, ±1.5%
K = 0.1 cm-1
K = 0.01 cm-1
K = 0.475 cm-1
K = 0.778 cm-1
Resolution
Depends on measuring range
Temperature sensor
Integrated NTC
Temperature Measurement
0 °C ... +130 °C
0 °C ... +100 °C
-5 °C ... 80 °C
0 °C ... 60 °C
Maximum pressure
14 bar (at 20 °C)
10 bar (at 20 °C)
2 bar
Electrical Connection
Integrated PU connection cable with 7-pole screw plug (IP 65)
Integrated cable mit 8-pole plug
8-pole socket for cable KKDU 325
Mechanical
Shaft
Stainless steel 1.4571
Epoxy
POM
Kable gland
Brass, nickel-plated
—
Connection head
Electrodes
Graphite
Protection Rating
IP65
IP68 (Sensor and cable)
IP65 (when connected)
(when connected)
(Measuring cell until immersion length)
Weight (without cable)
Approx. 350 g
Approx. 300 g
Approx. 135 g
Approx. 280 g
Approx. 170 g
Warranty
2 years on defects in quality according to § 10 terms of conditions
Data sheet D3.06, English, 999264USAnalog conductivity measuring cells
Data sheet D3.06, German, 999264DAnaloge Leitfähigkeitsmesszellen
Go to cart/checkout