• WTW PB-S Stationary samplers perspective view (right)
  • WTW PB-S/R1 Stationary sampler (503225), detail 1x25L
  • WTW PB-S Stationary samplers, detail view of control unit
  • WTW PB-S Stationary samplers frontal view, closed
  • WTW PB-S Stationary samplers frontal view of left side
  • WTW PB-S Stationary samplers frontal view of back side
  • WTW PB-S Stationary samplers frontal view of right side
  • WTW PB-S Stationary samplers perspective view (left)
  • WTW PB-S Stationary samplers perspective view (left), open top

Stationary sampler with vacuum technology

  • SKU: 503225
Request a Quote
  • Description

    Stationary sampler - PB-S with vacuum system for reliable, accurate and temperature-controlled sampling.

    • Stable sampling due to autarkic cooling or heating
    • Time-, volume-, event-, flow-proportional or manual sampling
    • Highly accurate single sample volume with vacuum system
    • Clear, simple operating structure and programming

    Due to the robust stainless steel housing, the samplers are particularly resistant and take your important samples in any weather. So nothing stands in the way of standard-compliant sampling.

    Models:

    PB-S/R1: Sampler 230V 50/60 Hz, with stainless steel housing and 1 x 52.5 pt. (25 l) PE sample container
    PB-S/R12: Sampler 230V 50/60 Hz, with stainless steel housing and 12 x 6.1 pt. (2.9 l) PE sample container
    PB-S/R24: Sampler 230V 50/60 Hz, with stainless steel housing and 24 x 2.1 pt. (1 l) PE sample container

    For flow proportional sampling, two additional models are available: 
    New from Xylem Analytics PB-S/R12-v: Like PB-S/R12, with variable dosing unit
    PB-S/R24-v: Like PB-S/R24, with variable dosing unit

  • Specifications
    Icon: CE Product Certification for Xylem Analytics Products  detailed technical data see data sheet D5.04
    Model PB-S/R1 PB-S/R12 PB-S/R12-v PB-S/R24 PB-S/R24-v
    Order-Nr. 503225 503228 503229 503230 503231
    Sampling fractioning 1 x 52.5 pt. (25 l) 12 x 6.1 pt. (2,9 l) 24 x 2.1 pt. (1 l)
    Sampling modes Time-, amount-, eventdependent or manual
    PB-S/R12-v and PB-S/R24-v: also flowproportional
    Dosing 0.042 ... 0.735 p. (20 ... 350 ml)
    Volume accuracy < 2,5 % or ± 0.0063 pt. (3 ml)
    Suction height Max. 26,25 ft. (8 m) at 1013hPa
    Programming 12 programs (freely programmable)
    Power supply 230V 50/60 Hz
    Status messages 1 x collective fault signal
    Data logging 3000 entries
    Interfaces Mini-USB, RS232

Downloads