ISE sensor for ammonium - WTW

ISE sensor for ammonium - WTW

  • SKU: 107072
Request a Quote
  • Description

    Digital ISE sensor for ammonium, connectable to the IQ SensorNet

    AmmoLyt®Plus 700 IQ

    Ammonium measurement directly in the medium without sample preparation and sample transfer. Measurement of centrate and other process waters up to 2,000 mg/l NH4-N:

    • Calibration free, long-term stability
    • No chemicals
    • Short response times

    The ISE sensor AmmoLytPlus 700 IQ allows ammonium measurement directly in the process. Time-consuming steps such as sampling and preparation are eliminated. For stable measured values, the interference substance potassium is compensated simultaneously and can be displayed if needed. An energy-optimized operation by a suitable regulation of the blower aggregates is made possible by this ISE sensor. This will help effectively reduce operating costs of the system, which is amortized by the system in a very short time. Starting now, you will be able to reliably determine centrate and other process water with high ammonium load up to 2.000 mg/l NH4-N with the AmmoLytPlus!


    Models:

    AmmoLytPlus 700 IQ: Digital probe for ion selective measurement of ammonium
    AmmoLytPlus SET/Comp: Sets, including Probe, reference electrode, measuring and compensation electrode 

  • Specifications

    Measuring method

    Electrochemical

    Appropriate Electrode

    Reference electrode VARiON® Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl

    Measuring range/Resolution

    NH4-N: 1 ... 2,000 mg/l / 1 mg/l; 0.1 ... 100 mg/l / 0.1 mg/l
    NH4+: 1 ... 2,580 mg/l / 1 mg/l; 0.1 ... 129.0 mg/l / 0.1 mg/l

    Compensation Range

    K+: 0.1 ... 1,000 mg/l / 0.1 mg/l

    Measuring Accuracy in laboratory standard solutions

    ± 5 % of measured value ± 0.2 mg/l in standard solutions

    Calibration Procedures

    Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution

    Working Life (typically)

    Reference electrode: 18 months,
    measuring and compensation electrode: 18 months (in typical application - municipal sewage plants)

    Temperature Measurement and Compensation

    Integrated NTC thermistor, 
    Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s

    Pressure Resistance

    Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes)

    Ambient Conditions

    Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C)

    Electrical connections

    2-wire shield cable with quick fastener to sensor

    Electromagnetic Compatibility

    EN 61326, Class B, FCC Class A;
    Intended for indispensable operation

    Certifications

    CE

    Mechanical

    Sensor body: V4A stainless steel 1.4571 Protective cup: POM
    Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes)
    Electrode connector: POM

    Weight

    Approx. 1.48 lb (670 g, without electrode, without sensor connection cable)

    Warranty

    AmmoLyt®Plus 700 IQ: 2 years
    Electrodes: 1 year for defects of quality

  • Accessories
  • Video

    Cleaning of ISE electrodes easy and simple

    Our robust ISE-electrodes enable a fast and easy cleaning with normal brushes (e.g. tooth brushes).